The Area Dose Rate Meter BD-01 is a hand-held dosimeter for
radiation protection measurements during application of
beta-nuclides.
Fields of application:
- Medicine, life-sciences, industry
- Local dose measurement in mixed beta-gamma radiation
fields
- Evaluation of workstations
- Retrospective dose assessments
- Localisation of beta-radiation sources and contamination
Download: BD-01 Data Sheet (*.pdf, 111 KB)
Poster-DASA 2003 (*.pdf, 187 kB)
Technical Specifications BD-01
Survey Meters for
Beta Radiation Type BD-01
Measureing
values: |
Ambient dose rate
equivalent dH*(10)/dt |
|
Directional dose
rate equivalent dH'(0,07)/dt |
Measurement range:
|
20 µSv/h to 200
mSv/h (standard version)
200 µSv/h to 2000 mSv/h (special version BD-01 HD)
2 µSv/h to 20 mSv/h (special version BD-01 ND)
|
Energy
range: |
Photons: 5 keV to
3 MeV |
|
Beta radiation: 80
keV to 3 MeV |
|
(Calibration using
beta-fields according to ISO 6980) |
Components: |
Detachable
measurement probe |
Detector: |
Ionisation chamber
(volume: 100 cm³) |
|
Shielding
material: Al and PMMA (550 mg·cm²) |
|
Entry window:
Al-deposited PETP-foil, (7 mg·cm²), |
|
Entry window can
be closed by a shield and a protection grid |
Display
Unit: |
- Numeric
Display
- Auto-Ranging
- Connected to
the measuring probe by a plugable cable
-
Battery-powered
- Manual
zero-adjustment possible
|
Weight: |
Display unit: 800
g (incl. batteries) |
|
Measuring probe:
400 g |
Dimensions: |
|
Display unit (L
x W x W): |
220 mm x 100 mm x
40 mm |
Probe (Ø x H x L): |
60 mm x 50 mm x
220 mm (handle: 50 mm) |