Measured
values: |
Photon equivalent
dose Hx |
|
Photon equivalent
dose rate dHx / dt |
Display
ranges: |
|
Dose rate: |
0 .. 2000 µSv/h
0 .. 2000 mSv/h |
Dose:
|
0 .. 2000 µSv |
Energy
range: |
6 keV to 7,5 MeV (approx. 15 MeV with additional
acrylic plastic shielding) |
Radiation
detector: |
Air-equivalent
ionisation chamber |
Volume: |
600 cm3 |
Build-up
cap: |
removeable, 550
mg/cm² |
Intrinsic
error: |
< 15 % (fine
measurement range 20)
< 5 % (fine measurement ranges 200 and 2000) |
Linearity: |
5% |
Minimum
Measurable Dose |
0,01 µSv/h |
Saturation deficite
: |
5% @ 2000 mSv/h |
Operating
voltage control: |
Battery symbol in
case of low battery voltage |
Interface: |
USB (optionally
with software) |
Power
supply: |
4 round batteries
LR 06 type AA max. 6,2 V
6V DC external power supply optionally available |
Power
consumption: |
approx. 30 mA @ 6V |
Battery
operating life: |
approx. 100 hrs |
Working
temperature range: |
-10 °C to + 45 °C |
Storage
temperature range: |
-25 °C to + 55 °C |
Ambient
pressure: |
800 hPa to 1100
hPa |
Relative
humidity: |
< 80 % |
Dimensions: |
|
Display unit (L x
W x H): |
250 mm x 108 mm x
42 mm |
Measuring probe
(Ø x H). |
112 mm x 260 mm |
Connecting cable: |
0,7 m (up to 100 m
optionally available) |
Weight: |
Display unit with
measuring probe: 1,5 kg |
|
|